CMS5000

瑞士INFICON VOC监测器 CMS5000

品牌名称:瑞士INFICON

产品型号:CMS5000

产品规格
产品介绍
洁净的水是生命必需的物质。由于环境中会有许多化学物质,因此对水质和相应健康影响的不利后果也会增加。为了达到最高的安保等级,需要快速响应水质问题。
渠道商
产品详情

产品介绍

洁净的水是生命必需的物质。由于环境中会有许多化学物质,因此对水质和相应健康影响的不利后果也会增加。为了达到最高的安保等级,需要快速响应水质问题。
基于 GC 的系统 CMS5000 能够自动提取水中的 VOC 样品,量化污染物水平,如果超过阈值将自动提醒用户。

 

产品优势

  • 通过自主测试节省人力和金钱

  • 减少常规外部实验室分析的频率

  • 最低的拥有成本

  • 减少时间浪费,较低的维护需求,无需准备样品

  • 易于集成,占地面积小

  • 分析更大范围的 EPA 8260 化合物

  • 能够在非常低的水平检测 VOC(ppt)

  • 可定制的方案能够满足您的特定需求

 

产品应用

  • 饮用水分配系统

  • 水源保护

  • 符合废水排放法规

  • 地下水污染整治

  • 空气质量监测

 

技术参数

Operating   Specifications


Power   requirement

120-240   V(ac)

Weight

55.1   lbs (25Kg) without sampling vessel

Dimensions

16.9"   x 32.7" x 10.2" (43 cm x 83 cm x 26 cm)

Operating   temperature

5-45°C;   95% RH (non-condensing)

Ionization   source

2.4   mCi Ni-63 Micro Argon Ionization Detector (MAID)

Storage

If   the CMS5000 is installed outdoors, it should be mounted under a covered   shelter. The unit should not be exposed to direct sunlight or driving rain.

Regulator   specification

M3300   regulator with CGA, quick connect, needle valve, pre-set to 90 psi  (IPN 935-412-P1)

Instrument   Information / Ship Kit


Ship   kit contents (IPN 935-721-GX)

Power   supply (IPN 930-469-PX)

Ethernet   cable (IPN 600-1319-P2)

Operator   manual (IPN 074-5021-G1)

Argon   tank regulator (IPN 935-412-P1)

Argon   fill line (IPN 935-212-G1)

CMS5000   training CD (IPN 074-5020-G1)

Quantitative   methods

Water   Purge Method (19 Compound VOC mix)

Check   Standard (Toluene Permeation Tube)

Please   contact INFICON for information regarding other available methods.

Sampling   vessel

The   CMS5000 can be used to monitor air or water. In order to switch between   sampling configurations, a modification will need to be made to the sampling   vessel (see  operator manual for   instructions). The optional water sampling vessel (IPN 935-700-G3) and the   air sampling vessel (IPN 935-701-G1) are available from INFICON.

Exhaust   lines

To   prevent backpressure, the two exhaust lines on the bottom of the instrument   should not be obstructed. One of the lines carries the exhaust from the   sample and the other carries the exhaust from the detector. When facing the   front of the instrument, the  sample   exhaust is on the left and the detector exhaust is on the right.

Installation   / Set-up


Securing   the instrument to a structure

CMS5000   is shipped with four (4) mounting brackets to support the instrument. The   mounting surface must be able to support 55.1 lbs (25 Kg), plus the weight of   the water vessel or air sampling vessel.    The total weight depends on what type of device is connected to the   sampling inlet on the instrument.

Minimum   mounting clearance

The   necessary space will depend on what type of sample vessel you are using. As   a  general rule, there should be   approximately 3’ of clearance between the bottom of the instrument and the   ground. The sample vessel must be able to be affixed to the chassis of the   CMS5000 allowing it to suspend freely without obstruction.

Line   input for water vessel

Line   input to the water vessel is ¼” NPT (National Pipe Thread) and output is 1¼”   NPT. The CMS5000 water vessel was designed as an open system to prevent water   from  getting into the sample path in   the event of a water vessel overflow. The output should be put into a waste   drain.

Required   flow rate through the water vessel

1   liter/min

Integration

TCP/IP   based; USB for local diagnostics, 19 pin I/0 relay contacts

Recommended   tubing for water monitoring

Copper   or stainless steel tubing is recommended. Polymer tubing is acceptable,   provided there is a constant flow of water to prevent VOCs from building up   in the pathway.

Recommended   tubing for air monitoring

Copper   or stainless steel tubing.

Factory   acceptance protocol

Water   analysis: Water Purge Method (19 Compound VOC mix) Air analysis: BTEX   standard

Consumables   / Maintenance

Carrier   gas requirement

Ultra-high   purity Argon (99.999%)

Carrier   gas line

Stainless   steel

Consumables

Tri-Bed   concentrator (IPN 930-716-G1)

Argon   gas (Sourced from your local gas distributor)

GC   column (HP-1, 0.32mm id, 30M,4.0 μm df or equivalent) (IPN 930-489-G14)   Toluene Permeation Tube (IPN 935-213-G1)

Check   standard

Toluene   Permeation Tube

Maintenance

The   water vessel and the air sampling tube must be kept free of debris.

Toluene   permeation tube will need to be replaced approximately every 8 years.

Argon   tank will need to be replaced as needed.

 