Mikron M390 超快速黑体校准源
Mikron® M390 黑体校准源提供行业领先的温度升温速率、温度范围和目标发射率。M390 标准 25 mm (1 in) 石墨腔靶管,在 0.65 至 1.8 μm 范围内的有效发射率为 1.0。M390 目标温度由快速响应红外测温仪感应,并由 PID 控制器调节,将目标温度保持在精确水平。M390 控制器安装在校准源上,可通过标准 RS232 通信端口进行远程设定点编程。
产品特点
PID控制器将目标温度精确调节到所需值
通过 RS232 进行远程校准设定点控制
读数的 ±0.25% ±1°C
在 0.65 至 1.8 μm 之间具有 1.0 的高有效发射率
配备可追溯至NIST的校准认证
在所需温度下快速稳定
实现高精度和可重复性
快速预热和冷却 — 从环境温度到 2300°C 仅需 5 分钟
轻松访问所有安全功能
满足严格的质量控制标准
Mikron M390 技术参数
型号 | Mikron M390 |
Measurement Specifications | |
Temperature Range1 | M390-A2 – PN 14029-A2: 600 to 2300°C (1112 to 4172°F) |
M390-B2 – PN 14029-B2: 600 to 2600°C (1112 to 4712°F) | |
M390-C2 – PN 14029-C2: 600 to 3000°C (1112 to 5432°F) | |
M390-L1 – PN 14029-L1: 300 to 2000°C (572 to 3632°F) | |
Temperature Uncertainty2 | ±0.25% of reading ±1°C |
Temperature Resolution | 1°C |
Stability3 | ±1°C |
Source Non-Uniformity | Non-uniformity: ±0.1% ± 1°C of reading within the center ½ of diameter |
Heated Cavity Shape | Closed end graphite tube 292 mm (11.5”) |
Exit Port Diameter | 25 mm (1”) |
Emissivity ε | 1.0 (effective from 0.6 to 1.8 μm) |
~0.96 (effective from 3 to 15 μm) | |
Calibration Method | Standard: Radiometric (pyrometric) |
Temperature Sensor | Custom infrared thermometer Impac IGA 6 |
Warm-up Time | 5 minutes from 600 to 2300°C, typical |
Slew Rate to 1°C Stability | ~325°C / min 600 to 1100°C |
~700°C / min T > 1100°C |
Communication and Electrical Specifications | |
Remote Set Point | RS232 serial communication output standard, RS485 optional |
Method of Control | Digital PID controller |
Power Requirements | M390-A2 – PN 14029-A2: 15k VA 1 ph |
M390-B2 – PN 14029-B2: 15k VA 1 ph | |
M390-C2 – PN 14029-C2: 24k VA 1 ph | |
M390-L1 – PN 14029-L1: 15k VA 1 ph | |
208 to 230 VAC @ 50 and 60 Hz, 15k VA or 24k VA (depends on model) | |
Power Connector (Customer Supplied)1 | 15k VA: Must use plug that complies with IEC-309 @ 75 Amps @ 220 / 230 VAC |
24k VA: Must use plug that complies with IEC-309 @ 125 Amps @ 220 / 230 VAC |
Environmental Specifications | |
Operating Ambient Temp | 0 to 44°C (32 to 110°F) |
Cooling | Water cooled. Standard (garden) hose connections on rear, |
Cooling Inlet Pressure | 90 PSI (621 kPa) maximum. Customer is responsible to provide pressure safety |
Purge Gas | Argon High Purity (99.99% pure) |
Purge Gas Inlet Pressure | 25 PSI (172 kPa) maximum. Customer is responsible to provide over pressure safety valves at inlet |
Heating Element Type | Graphite tube with argon gas purge, 226 L/h (8 to 10 cfh) |
Operating Humidity | 0 to 90% RH non-condensing |
Dimensions (H x W x D) | 1710 x 560 x 820 mm (67.3 x 22 x 32.3 in) |
Weight | 182 kg (400 lb) |